Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

In rehab, the ECG signal is evaluated for its quality. The program analyses the signal quality separately for each ECG channel and assigns one of the following signal qualities every second:

ECG signal qualityExample ECG signalScore value in information blockDescription
good ECG quality

Image Modified

E:1ECG must look like this, with such an ECG signal there are guaranteed no problems with the QRS detection
sufficient ECG quality

Image Modified

E:3This ECG quality means that the QRS detection will still recognize every QRS complex due to internal filtering, but it is already an indication that one or the other QRS complex may be overlooked here.
insufficient ECG quality

Image Modified

E:6This ECG quality means that the QRS detection is certainly "overlooked" despite the internal filtering of some QRS complexes (false negative) or disturbances of the ECG signal even as QRS detection (false positive).
non-physiological ECG quality

Image Modified

E: !!With such an ECG Signal no QRS complex can be detected. This occurs, for example, when the electrodes of the ECG conduction are open.


The ECG signal quality information is used for the following actions/displays

...

The visibilty of ECG signal quality can be activated in Option 4.13 under  Settings → Reha → General → Menu/Functions → Additiona Additional settings for exercise stations → Display options un in the exercise.

                   

...

In case of a bad ECG signal quality a transparent red square is faded fade in over the ECG in the online ECG part. With a 3 channel ECG , the rectangle is vertically divided into three equal areas, one . One for each channel. This allows the quality of all three channels to be signalled simultaneously even if the currently displayed ECG channel is not affected.

...

In contrast to the graphic quality display, the numerical displays are always retained, even if the ECG is of good quality. 

As

...

score value

Any information block cann can be configured to disaply display the ECG quality score value


1 channel ECG

Image Modified

The ECG channel has a quality score of 6, which means insufficient ECG quality.
3 channel ECG

Image Modified

In this three-channel ECG display, the ECG channel 1 is of poor quality (score vauel of 6) and the other two channels are not physiological (score value of 20). 


As numerical vaue

The determination of the ECG signal quality has as result a numerical value, which can be interpreted as a measure of the noise. This can be displayed as a single numerical value for each channel. The value itself is more of a technical specification and should therefore only be displayed in disputed quality cases. 

...