Introduction:

The article discribes the Workflow and the configuration from the file-Interface for cardio pulmonary exercise Testing.

Custo diagnostic is always master and the Spirosystem slave. That means: This spirosystem can only be controlled by custo diagnostic!!

Manufacturer which support this interface

Viasys

SensorMedics

Jäger

CareFusion

Ganshorn

Connection about shared File "ergospir.dat"

diag writes the  parameters every 5 seconds during the spiroergometry exam Text-file c:\\temp\ergospir.dat.

The path is configurable.

Configuration diag

diag > Settings > System (1)> System (2)> Workstation (3) >cardio pulmonary exercise Testing = for example: Ganshorn (LF8.5G) (4)


 

file "ergospir.dat"

diag writes a text file every 5 seconds. with the current ergometry data.

Path

c:\\temp\ergospir.dat

 

The path is configurable in the win.ini file in the Windows directory:

[CAS_RECORD]

SpiroDirectory=C:\\Temp

Content of the file

Example file:

View file
nameErgospir.dat
height250

[exer   25 W    0 %   75    0 -999 -999 -999  -12 -999 -999]11232

 

Introduktion:

  1. always an open square bracket
  2. Phase, one of the following strings (in brackets stands the meaning):
    chck
    base
    warm (warm up)
    exer (start examination)
    reco (cool down)
    stop
  3. load values
    by Treadmill:
    Speed [km/h]
    Sign: K
    Inclination
    Sign: %
    by Ergometer:
    Load
    Sign: W
    Sign: 0
    Sign: %
  4. Heart rate
  5. always 0
  6. always-999
  7. Systole, if not available -999
  8. Diastole, if not available -999
  9. ST-Value
  10. always -999
  11. alwaya -999
  12. always a closed square bracket
  13. Checksum from the previous values

...


Workflow for Start from cardio pulmonary exercise Testing :


  1. Start the Spirosystem firstly and then 
  2. Start custo diagnostic Project: "cardio pulmonary exercise Testing."
  3. available the Patient, Profile and click start 
  4. When switching to the Monitoring page in custo diagnostic, the patient data become now transferred to the Spirosystem
  5. Start the exercise Testing in custo diagnostic