Table of Contents |
---|
Introduction
The article discribes the Workflow and the configuration from of the file -Interface based interface from certain manufacturers (see list), for cardio pulmonary exercise Cardio Pulmonary Exercise Testing.
Custo custo diagnostic is always the master and the Spirosystem Cardio Pulmonary Exercise System is slave. That means: This spirosystem Cardio Pulmonary Exercise System can only be controlled by custo diagnostic!!
Systems which support this interface
- Viasys
- SensorMedics
- Jäger
- CareFusion
- Ganshorn
Connection through shared File "ergospir.dat"
custo diag iagnostic writes the parameters the parameters every 5 seconds during the spiroergometry exam Text-file c:\\temp\the Cardio Pulmonary Exercise Test to the text file "ergospir.dat".
The path is configurable.
Configuration diag
diag > Settings > System (1)> System (2)> Workstation (3) >cardio pulmonary exercise Testing = for example: Ganshorn (LF8.5G) (4)
file "ergospir.dat"
diag writes a text file every 5 seconds. with the current ergometry data.
Path
to this file can be configured.
Path
Default: C:\Tempc:\\temp\ergospir.dat
The path is configurable can be configured in the file "win.ini" file in the Windows directory:Directory:
You have to add this Section:
Code Block |
---|
[CAS_RECORD] |
SpiroDirectory=C:\\Temp |
Configuration custo diagnostic
- Start custo diagnostic
- Go to Settings
- Go to System (1)
- Click on System (2)
- Click on Workstation (3)
- Select th apropriate Interface at Cardiopulmonary Exercise Testing (for example: Ganshorn (LF8.5G)) (4)
Workflow to Run a Cardio Pulmonary Exercise Testing (Start from custo diagnostic)
- Start the Cardio Pulmonary Exercise System first
- Start custo diagnostic
- Open "Cardiopulmonary Exercise Testing"
- Select the Patient
- Select the Testing Profile
- Click "Start"
Workflow for Start from cardio pulmonary exercise Testing
- Start the Spirosystem first and then
- Start custo diagnostic Project: "cardio pulmonary exercise Testing."
- select the Patient, Profile and click start
- When switching to the Monitoring page in custo diagnostic, the patient data become now will be transferred to the SpirosystemCardio Pulmonary Exercise System
- Start the exercise Cardio Pulmonary Exercise Testing in custo diagnostic