Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

From custo diagnostic version 4.6.11, 5.0.7, 5.1.1 and 5.2.0, custo med built in an additional feature that gives you the possibility to adjust the sensitivity of the QRS trigger output. ( The QRS trigger output is an option that needs a QRS Trigger cable and an update of the license, in case the end-user wouldn't have it already)

Under normal conditions, what means the ECG has an ECG with a average amplitude and not much noise or no noise on the ECG, the QRS trigger works properly. In some occasions when the amplitude of the ECG is very small the trigger might every now and then miss a R top. In case of using the trigger pulse for a CT camera or Blood pressure device this might affect the measurement.

...