Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Fieldo/rDefault Usage



PV1r

Visit-ID. The visit id is used to identify the visit in custo diagnostic It must be unique in one custo diagnostic installation, otherwise an appropriate tenant system must be used (see tenant system below).

PV1-44oAdmission Date: custo diagnostic can store this information, even it is not shown in the user interface.  Its done for future usage.
PV1-45oDischarge Date/Time: custo diagnostic can store the discharge date/time to filter out discharged patients in patient search lists. It can be defined, how custo diagnostic interpretes an empty discharge date/time - either als missing, unknown information, or as the information, that the discharge date should be deleted in custo diagnostic. 



ORC - Segment


Fieldo/rDefault Usage

r

The combination of OBR-1 and OBR-5 can be configured as order control:



ORC-16oORC-16-2: Order Control Code Reason / Clinical questions, shown to the user before processing the observation request.



OBR - Segment


## bitte hier alles aus default mapping nehmen##

...

Fieldo/rDefault Usage
OBR-1 / OBR-5r

The combination of OBR-1 and OBR-5 can be configured as order control:



OBR-2rPlacer ID
OBR-3oSending: Outgoing:  Consecutive Number with an alphanumeric prefix, e.g. "CM00001234" or some GUID, e.g. "36eea93b-bc1b-4769-a84e-4ebeaaaf387e"
OBR-4r

OBR-4 contains information about the type of observation which should by executed.

OBR-4-1 is a coded value which must be defined.

OBR-4-2 when uutgoing, OBR-4-2 is shown to the user, e.g. "Resting ECG"

OBR-4-1 and OBR-4-2 is requried, other components optional. 

##bitte prüfen - when OBR-4-2 in eingehender Nachricht abweichen zur Konfiguration in "Untersuchung" - welcher Text wird dabei angezeigt - der aus der Konfig, oder der aus der Nachricht?




OBR-20oIncoming: Flexible text-fields which can be used to show additional information regarding this order to the user.
OBR-21oIncoming: Flexible text-fields which can be used to show additional information regarding this order to the user.



Patient Data Logic (ADT)


Patient / Visit Identification

...

custo diagnostic receives orders and show them as a worklist to the users. One order is (by default) identified by the Placer Number, and may contain serveral several observation requests, identified by Alternate Order ID. A clinical question is visible to the user, which can contain a free text and to avoid communication by E-Mail/Phone etc. 
The user choose the requested observation, and custo diagnostic determines the diagnostic mode by the service identified given in OBR-4, e.g. Spirometry or Resting-ECG.

When the orders are processed, custo diagnostic adds an internal "Filler ID" . This can be configured by a fixed length consecutive number, with an arbitrary alphanumeric prefix, e.g. "CM00001234", or an GUID.



Observation Results (ORU)

...