Field | o/r/i | Default Usage | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
OBR-1 / OBR-5 | r | Order Control The combination of OBR-1 and OBR-5 can be configured as order control: Default:
By default, OBR-5 is not evaluated, but can be configured to specific values. In this case the combination of OBR-1 and OBR-5 must match. ORR-Messages: See below, "Updating order placer via ORR" | |||||||||||||||
OBR-2 | r | Placer ID | |||||||||||||||
OBR-3 | o | Outgoing: Consecutive Number with an alphanumeric prefix, e.g. "CM00001234" or some GUID, e.g. "36eea93b-bc1b-4769-a84e-4ebeaaaf387e" | |||||||||||||||
OBR-4 | r | OBR-4 contains information about the type of observation which should by executed. OBR-4-1 is a coded value which must be defined. OBR-4-2 when incomfing, OBR-4-2 is shown to the user, e.g. "Resting ECG" OBR-4-1 and OBR-4-2 is requried, other components optional. ##bitte prüfen - when OBR-4-2 in eingehender Nachricht abweichen zur Konfiguration in "Untersuchung" - welcher Text wird dabei angezeigt - der aus der Konfig, oder der aus der Nachricht? | |||||||||||||||
OBR-5 | see OBR-1 | ||||||||||||||||
OBR-20 | o | Incoming: Flexible text-fields which can be used to show additional information regarding this order to the user. | |||||||||||||||
OBR-21 | o | Incoming: Flexible text-fields which can be used to show additional information regarding this order to the user. |
...