Consider starting custo diagnostic in general with the extended search and always check for the relevant evaluation.



I

...

get many evaluations and I don't want to wait for the initial analysis - can the initial analysis be automated?

Yes, via the task manager of the custo diagnostic client.

In the settings of the client set diagserver.datatransfer.upload.createAnalysisJob to TRUE.

Then enable the automatic execution of tasks:

Image Added

The sender may have sent the evaluations for a patient several times. This can happen by accident, by not trusting the sending mechanism or because the sender has modified the evaluation (e.g. the sender may have added some comments in the report field), So never just delete an evaluation - first communicate in this case with the sending practice about the background. If you sure that you can delete duplicate evaluations select the relevant patient, go to the patient menu, click on "Edit Database" and then on "Delete Evaluations". mark the duplicate evaluations and click on "Delete Marked Evaluation".

The hard disk of my evaluation center gets filled up only a few a hours after the custo diagnostic server has started

With the default setup paths it is the C:\Program Files\custo diagnostic server\temp directory which shows many many temporary files. A workaround is to stop the custo diagnostic server. With the next custo diagnostic server start the files will get deleted. A final solution is to upgrade the custo diagnostic server to version 5.3.4 or 5.4.1.

sending GP


Communication to the evaluation timed out / no connection

...

Looking at the RDT tool log you may see entries like "unable to unzip file...". Verify that the RDT tool has read and write access to the data directories you have defined for the incoming and outgoing data.


What is the difference between the RDT tool 2.0 and 1.3?

The RDT tool 2.0's intended use is for version 5 evaluation centers to there simulate a custo diagnostic version 4 evaluation center.

Image Added

Multiple channels can be defined there with on the one side file import / export directories for each sending practice and on the other side the corresponding RDT connections.

The RDT tool 1.3 is intended to simulate a version 5 sending practice on a computer vwhere only diag version 4.x is installed.