custo diagnostic is a server based software (referenced as "software"), intended to support physicians to acquired, analyze, document and report cardio pulmonary observations. The inherent patient, order and observation database can be interfaced via HL7, DICOM and GDT to an superior system. This document describes the typical use of the HL7 interface version 5.6.4 and above and addresses the HL7 communication partner. (The intention is not, to give a deep insight about the configuration in our service applications to configure this options). 

...

i - ignored. 


General HL7 information


Supported HL7 Version2.3, 2.5 (global for all message-types)
Character Sets

All typical character sets supported.

In general, custo diagnostic is configured to export via Windows-1252 (global for all export messages)

Supported Transfer 

TCP/IP Socket

File based

Supported Message Types
  • ADT
  • ORM
  • ORR
  • ORU
  • DFT
  • MDM
  • ACK

The message types must be licenced individually - which means not all installations provide all message types.

Empty Value /
Null Value

In general custo diagnostic supports the null value for receiving messages.



Interpretation
||The value is unknown. A corresponding data field in custo diagnostic is not changed
|""|The value is empty. A corresponding data field in custo diagnostic will be cleared.



HL7 Segments

This HL7-Segements section describes the typical mapped used data fields and components . For special purpose and installations, a divergent mapping can be agreed- it is not a complete mapping list, because it should focus on a reasonable usage of the system.

In case more information should be transferred or processed - please refer to the authorized system vendor




Optionality Key

Incoming

...

This means, when receiving a message MSH-15 / MSH-16 is ignored, because the general configuration is taken. When sending a message, MSH-15 will depict the configuration in custo diagnostic. MSH-16 will always be "NE".

PID - Segment

### Hier bitte alles aus default Mapping übernehmen.###

...

Required sind patienten name/vorname, geburtsdatum?, Patienten-nummer, geschlecht?


Fieldo/r/is/soDefault Usage
PID-3rs

PID-3-1 is used

...

to identify the patient in custo diagnostic.

It must be unique in one custo diagnostic installation, otherwise an appropriate tenant system must be used (see tenant system below).

Required sind patienten name/vorname, geburtsdatum?, Patienten-nummer, geschlecht?

PID-3-1 is used to identify the patient in custo diagnostic. It must be unique in one custo diagnostic installation, otherwise an appropriate tenant system must be used (see tenant system below).
PID-5rs

PID-5-1, PID-5-2, PID-5-3, PID-5-4, PID-5-6:  Family Name, Given Name, Second Name, Suffic, Degree/Title 

Required: PID-5-1

PID-7rsDate of birth
Fieldo/r/is/soDefault Usage
PID-3rs
PID-8r ?s

Administrative Sex. Supported Values are F, M, ?? Other Values are interpeted as ... 

##zu prüfen##


PID-11osAddress of the patient


PV1 - Segment


##bitte aus default mapping ergänzen## Ist wahrscheinlich alles optional

...