Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Patient Data Logic (ADT)

ADT Messages

custo diagnostic accepts by default the following message types (further message types can be defined):

Supported messages:  A01, A02, A03, A04, A05, A08, A11, A12, A13, A23, A29, A38, A40, A42, A43, A45 (further message types can be defined)

Patient / Visit Identification

...

Tenant System

Tenant Systems (german "Mandant") can be configured in different ways to separate orders/observation and/or patients/visits. 

...

Tenant (Description)

Receiving (e.g. ORM)

MSH-6-1 (code)

Sending (e.g. ORU)

MSH-4-1

Sea View Hospital
SVHSVH
Lake Hospital
LHLH

...


Order Data / Order Entry Workflow

...

1) Incoming: See section "Orders" above.



Sending specific observation data

...

This HL7-Segements section describes the typical used data fields and components - it is not a complete mapping list, because it should focus on a reasonable usage of the system.In case more information should be transferred or processed - please refer to the authorized system vendor.  

Components which are affected by the workflow are described in the above data handling / workflow sections.


Optionality Key

Incoming

rrequiredThis value is required for a proper function of the software
ooptionalThis value is not required but may be used for some functionality in custo diagnostic. 
iignoredThis value is ignored

...