Use a calibrated patient simulator to perform a signal with a frequency of 1 Hz that starts with a 1mV pulse. Start custo diagnostic and perform a resting ECG test. Disable all filters (Anti Drift filter, muscle filter, 50 Hz filter) in the settings. Measure the heart frequency and confirm the value is the same as set in the simulator.Measure the 1mV pulse at the limb channels (see screenshot). With a scaling of 10 mm/mV, the Confirm that the waveform is shown as in the screenshot.
4. Pressure leakage (only cardio2xx)
...
Appendix A: permitted values
Source: NEN-EN-IEC 60601-1 (as referenced by NEN-EN-IEC 62353)
...