The problem

You have started an recorder with an patient A about 24 hours. Normally this patient comes back after 24 hours and you will read in all the measurement values in the software.
But now, you will give further this recorder to a new patient B and you 've forgot to start the recoder again for this patient.
This is the correct way!.

What will happen?

If you download this evaluation from patient B, so you will get a mixded information about 2 differnt patients.
A and B !
This isn't your target!

Which settings you should be noted

Please use this link


When you got an mistakeIf you got

an recording with mixed measurement data of 2 differend patients you can Unfortunately you can't fix this issue, but you could do followings:


Image RemovedImage Added


  1. select the patient
  2. select his examination
  3. click on button "Table"
  4. click on button "Delete Measurement" - please decide one of the 2 possibilities



  5. or Disable an part of the measurement, which will not longer calculated total quantity!


How you can prevent this issue?

Please use this link. There you can find some settings.