Introduction

...

The article discribes the Workflow and the configuration

...

of the file

...

based interface from certain manufacturers (see list), for Cardio Pulmonary Exercise Testing.

...

custo diagnostic is

...

always the master and the

...

Cardio Pulmonary Exercise System is slave. That means: This

...

Cardio Pulmonary Exercise System can only be controlled by custo diagnostic

...

!

...

Systems which support this interface

  • Viasys
  • SensorMedics
  • Jäger
  • CareFusion
  • Ganshorn

...

Connection through shared File "ergospir.dat"

custo d

...

iagnostic writes

...

the parameters every 5 seconds during

...

The path is configurable in the win.ini file in the Windows directory:

the Cardio Pulmonary Exercise Test to the text file "ergospir.dat".
The path

...

Configuration diag

diag > Settings > System (1)> System (2)> Workstation (3) >cardio pulmonary exercise Testing = for example: Ganshorn (LF8.5G) (4)

Image Removed

 

file "ergospir.dat"

diag writes a text file every 5 seconds. with the current ergometry data.

Path

c:\\temp\ergospir.dat

 

to this file can be configured.

Path

Default: C:\Temp\ergospir.dat
The path can be configured in the file "win.ini" file in the Windows Directory. You have to add this Section:

Code Block
[CAS_RECORD]

...


SpiroDirectory=C:\\Temp

Content of the file

Example file:

View file
nameErgospir.dat
height250

[exer   25 W    0 %   75    0 -999 -999 -999  -12 -999 -999]11232

 

Introduktion:

  1. always an open square bracket
  2. Phase, one of the following strings (in brackets stands the meaning):
    chck
    base
    warm (warm up)
    exer (start examination)
    reco (cool down)
    stop
  3. load values
    by Treadmill:
    Speed [km/h]
    Sign: K
    Inclination
    Sign: %
    by Ergometer:
    Load
    Sign: W
    Sign: 0
    Sign: %
  4. Heart rate
  5. always 0
  6. always-999
  7. Systole, if not available -999
  8. Diastole, if not available -999
  9. ST-Value
  10. always -999
  11. alwaya -999
  12. always a closed square bracket
  13. Checksum from the previous values

Workflow:

...

Configuration custo diagnostic

  • Start custo diagnostic
  • Go to Settings
  • Go to System (1)
  • Click on System (2)
  • Click on Workstation (3)
  • Select th apropriate Interface at Cardiopulmonary Exercise Testing (for example: Ganshorn (LF8.5G)) (4)

Image Added

Workflow to Run a Cardio Pulmonary Exercise Testing (Start from custo diagnostic)

  1. Start the Cardio Pulmonary Exercise System first
  2. Start custo diagnostic
  3. Open "Cardiopulmonary Exercise Testing"
  4. Select the Patient
  5. Select the Testing Profile
  6. Click "Start"
  7. When switching to the Monitoring page in custo diagnostic, the patient data

...

  1. will be transferred to the

...

  1. Cardio Pulmonary Exercise System
  2. Start the

...

  1. Cardio Pulmonary Exercise Testing in custo diagnostic