The new option is named “Realtime-QRS-Trigger-Module” and can be ordered by using part number 12149/LODE ( this exists out of a cable and a software license)

 

The trigger cable that can be used to trigger the internal blood pressure device, must be used for the ergometers and treadmill out of the following range:

...

  • 960900 Corival CPET        starting from serial number 960900-20150001
  • 960901 Corival REHAB    starting from serial number 960901-20150001
  • 912900 OEM ergometer    starting from serial number 912900-20150001
  • 938900 Valiant CPET        starting from serial number 938900-20141001
  • 938901 Valiant REHAB    starting from serial number 938901-20141001
  • 939910 OEM treadmill       starting from serial number 939910-20151001
  • 967900 Angio CPET          starting from serial number 967900-20170001
  • 967901 Angio REHAB      starting from serial number 967901-20170001

     

     


The trigger cable has a DB-9 connector on one end and a 2,5 mm2 on the other end. The cable must be connected to the PC  using an RS-232 port. A serial to USB converter might cause a delay in the trigger pulse which will affect correct blood pressure measurements, We strongly recommend NOT to use USB to serial converters because of that.