Introduction

If an recorded ECG has massive disturbances so that the ECG lines will be printed outside of the visible area. The printout results like so:

Image Added

...

This Article describes the  Settings for a Correction of the amplitude size in the printout.

incorrect representation in the printout

Image Removed

Fix the problem

a Setting to clip the ECG lines which are outside of the visible area.

Solution

Open the ServiceCenter an go to the Settings Editor. Set the Setting set the Setting in the Servicecenter "Langzeit-EKG\Drucken\EcgExample\EcgClip" of  to "True".


Result:

Image Modified