The custo screen 300/400 ABPM Recorder can store Data up to 3 Days. After the Recorder was read in (Download Data) the Recorder is not reset, so the programmed Patient Data and previous stored data stay on the Recorder. If the Recorder is switched on after read in, the Recorder continues the recording. If now the Recorder is applicated on a new Patient, the Recorder stores the new measured Values to the Storage Set of the previous Patient.

To prevent this, custo diagnostic support the Function "Start the Recorder after Read in". "Start the Recorder" means "Delete the Data from the Recorder".

To activate this Function go to this Settings Page in ABPM:

Image Added

FunctionDescription
neverdo nothing after read in the Recorder
alwaysStart the Recorder after every read in (Delete the Data)
with enquiryAsk the Customer if the Recorder is to be started after read in

Image Removed

Click the button Step 1 - Step 3 in custodiagnostic please.

Step 4

  • "never"  - Generally is selected after installation of custodiagnostic

The measurement data will never delete, also not, if you downloaded all measurement data from the last patient! This will happen only then when you start the recorder with an new patient.

...