the EDAN SE-2012 is only supported till custo diagnostic version 5.8

Technical specifications


SpecificationsEDAN
Channel(s)12 channel ( using 10 leads)
Sample rate128/256/512/1024 Hz
Resolution8/12/14/16/180 Bit
Storage2 GB (using micro SD)
Record time24/48 hours
Pace maker recognition± 2mV ~ 200
Dimensions76 * 49 * 16 mm
Weight50 grams without battery
Material, housingPlastic
Power supply1 x triple A Battery (AAA)
Runtime24 h at 1024 Hz and 144 h at 128 Hz
Battery voltage1,2 ~ 1,6 Volt
Frequency response0,05 ~ 60 Hz at -3 dB
Working temperature+5 ºC till +45 ºC
Protection classIP 27
DisplayColor LCD
Reading DataSD-card or USB cable
LanguagesGerman, English a.o.
Compatibilitycusto diagnostic 4.6.3.1 ( License #50245)

...

Make sure that the first time you start the recorder the recorder is started with custo diagnostic. If you start the recorder for the first time without connection to custo diagnostic the license on the recorder is overwritten. You can make a copy of the license file on the SD card before starting the recorder. In case you have overwritten the license you can contact service@customed.de with the serial number of the EDAN SE-2012 they can create a new license for you.


In the service center can be checked if the EDAN Recorder license is activated.



Settings Menu

Below you find the default settings, these are the settings, set when the device leaves custo med GmbH.


Image Added Settings


Gain:                        1

Channels:               12

Record time:           1

ECG Preview:        On


Patient ID:            Off

ECG Display:         On

Data Protected:   On

Language:      English


System Time:

             XX-XX-XXXX

             xx:xx:xx

Date Format

                  D-M-YY


Battery Type:

             alkaline       

Exit




Image Added Advanced Settings

AD Resolution:     16

Sample Rate:      128

Other Settings:    Off

Upgrade

Exit


Image Added Device Info

Model:

             SE-2012        

Software:

                 V1.11

Hardware:

                  V1.10