Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

Problem

After read in the ABPM recorder you got a 48h mixed Evaluation from 2 different Patients.

What is happened

You The problem
You have started an recorder with an patient Patient A about 24 hours. Normally the recorder runs for 24h and after this patient comes back after 24 hours and you will read in all the measurement values in with the software.

But now, without programming (starting) the recorder with patient B you will give further this recorder to a new patient B. The patient B is wearing the recorder also for 24h and you 've forgot to start the recoder again for this patient.
This is the correct way!.
What will happen?
If you download this evaluation from patient B, so you will get a mixded information about 2 differnt patients.
A and B !
This isn't your target!

Solution
comes back to read in the measuring values by you.

Solve the Problem

To solve the problem you have to delete all measured values of the patient A from the evaluation of the patient B. You ca do this in the Measurement Values Table:

  • Open the Evaluation of patient B
  • go to the Table
  • mark the latest measurement of patient B
  • click the right mouse button and select "Delete Meas." to delete the Measurement:

Image Added

Than you get this Message Box and you have to decide either to delete only one measurement (Option 1) or to delete all previous measurements up to the selected (Option 2):

Image Added

Disable an part of the measurement, which will not longer calculated total quantity!

Image Added

How you can prevent this issue?

Please visit this Page.Please use this link about