Introduction:

The article describes, how can i activate an configure the maintenance intervall


Basic:

order the keyoption "Maintenance":

Configure the Maintenance Intervall:

  1. open the webepage https://www.customed.de/service/Maintenance-Code-diag-5x,603

(1) Click the Link "Maintenance Code diag 5.x" (2) Input the Keynumber and Click (3) "Check"


  2. generate the maintence code:


(1) Input the Expiration Date (2) available the Warning Level (see table maintenance level) (3) Click "create"

Maintenance Level

StepDefinitionIncrease to next level (when maintenance interval has expired)
0No notice, maintenance interval has not yet expiredNo automatic increase (note only in the info dialog)
1Notice when starting the programafter another 14 days
2How 1 and additionally when calling up a projectafter another 14 days
3Like 2 and additionally with every printout and exportafter another 30 days
xMaximum warning level according to the settingafter another 30 days (but has no further effects)


3. automatic key generation:


webpage with Maintenance Code

and automatic generate a E-Mail with the same information:


4. Input and activate the Maintenance Intervall in custo diagnostic


start custo diagnostic (1) System (2) System (3) Maintenance (4) Input the Maintenance Code (example from this page "F05D3F6C74") and Click of the keybord "Enter" (5) Result you can see "Next data" and the "Warning Level"