General
The ECG Monitoring can be used as Viewing Module for Customers who want to observe Resting or Stress Test ECG Procedures where the Viewing Place and the ECG Place is spatially separated.
Installation
The ECG Monitoring Module is integrated in custo manager.exe and must not be installed separately. With the Installer you can create a Shortcut on your Desktop to start custo manager.exe as the ECG Monitor Module.
Th Shortcut started with the command "C:\Program Files (x86)\custo diagnostic client\exe\custo manager.exe" /monitoring.
Using
To start the ECG Monitoring Window you have to double click the ECG Monitoring Shortcut on your Desktop.
In the next Step the custo diagnostic Logon will be displayed. You have to logon on custo diagnostic with your credentials.
After the logon th ECG Monitor Window will be displayed.
If any ECG Measuring Desk is sending ECG the Patient Name will be displayed in one of the Channel Buttons.
Settings
Normally the ECG Monitor Window is started with the full Monitor width. If you want to run the ECG Monitoring Window in a smaller Width you have to state the Command Line Option "/Res=" in the Shortcut.
All other Settings for the ECG Monitoring Window are separated saved in the Configuration Database in CSC->WindowedEcgStreaming.
Setting | Default Value | Description |
---|---|---|
NumberOfStripes | 1 | Numer simultanous displayed ECG Stripes. Can be set to 1 or 2. |
PatientButton1stLineSettings | 27 | Date of Birth of the Patient |
PatientButton2ndLineSettings | 42 | Name of the ECG Sender PC |
WindowPosX | 0 | last X Position of the Window |
WindowPosY | 0 | last Y Position of the Window |
WindowSizeY | 0 | last Y Size of the Window |
ECG-Stripe1\EcgChannel | 16 | selected ECG Channels |
ECG-Stripe1\Gain | 10 | ECG Amplitude (10 mm/mV) |
ECG-Stripe1\RrViewMode | 1 | Selection between HR or RR Time in the Line below the ECG |
ECG-Stripe1\WriteSpeed | 10 | ECG Speed (10 mm/s) |
ECG-Stripe2\EcgChannel | 16 | selected ECG Channels |
ECG-Stripe2\Gain | 10 | ECG Amplitude (10 mm/mV) |
ECG-Stripe2\RrViewMode | 1 | Selection between HR or RR Time in the Line below the ECG |
ECG-Stripe2\WriteSpeed | 10 | ECG Speed (10 mm/s) |