if you need to perform maintenance on the cardio 130 refer to the following information on recurrent tests on cardio 1xx

Repairs

There are a few repairs that, service personal trained by custo med, can perform themselves.

  • Exchanging the USB cable
  • Exchange the 15 pin sub D-connector

Exchanging the USB cable

The USB cable has part number 39.415.33


In case the USB cable must be replaced you can order the part with the above mentioned part number. The part will come with a socket that can be soldered on the circuit board. Cardio 130 devices manufactured before 2014 had a different socket on the circuit board. the supplied USB will not fit on that socket and in that case you have to exchange the socket as well.


Exchange the 15 pin sub D-connector

In case you need to exchange the Sub D-connector you have exchange part 39.411.34

If the black part in which the connector is mounted needs to be replaced as well you have to be aware that there are two models.

  • Model 1, part number 39.326.19 this has a Siemens click connector.
  • Model 2, part number 39.209.07 that can be fixed with two screws.


Pin Layout D-connector patient cable

Electrodes according American Heart Association & International Electrotechnical Commission

AHAIEC
PinAHAColour
 
1V2
C2

              

2V3
C3
3V4
C4
4V5
C5
5V6
C6
6SHIELD


7NC


8NC


9RA-(R)


10LA-(L)


11LL-(F)


12V1


13NC


14RL-(N)


15

NC