if you need to perform maintenance on the cardio 130 refer to the following information on recurrent tests on cardio 1xx
There are a few repairs that, service personal trained by custo med, can perform themselves.
The USB cable has part number 39.415.33
In case the USB cable must be replaced you can order the part with the above mentioned part number. The part will come with a socket that can be soldered on the circuit board. Cardio 130 devices manufactured before 2014 had a different socket on the circuit board. the supplied USB will not fit on that socket and in that case you have to exchange the socket as well.
In case you need to exchange the Sub D-connector you have exchange part 39.411.34
If the black part in which the connector is mounted needs to be replaced as well you have to be aware that there are two models.
Electrodes according American Heart Association & International Electrotechnical Commission | ||||
AHA | IEC | |||
Pin | Inscription | Colour | Inscription | Colour |
---|---|---|---|---|
1 | V2 | BROWN/YELLOW | C2 | WHITE/YELLOW |
2 | V3 | BROWN/GREEN | C3 | WHITE/GREEN |
3 | V4 | BROWN/BLUE | C4 | WHITE/BROWN |
4 | V5 | BROWN/ORANGE | C5 | WHITE/BLACK |
5 | V6 | BROWN/PURPLE | C6 | WHITE/VIOLET |
6 | SHIELD | - | SHIELD | - |
7 | NC | - | NC | - |
8 | NC | - | NC | - |
9 | RA | WHITE | R | RED |
10 | LA | BLACK | L | YELLOW |
11 | LL | RED | F | GREEN |
12 | V1 | BROWN/RED | V1 | WHITE/RED |
13 | NC | - | NC | - |
14 | RL | GREEN | N | BLACK |
15 | NC | - | NC | - |