Introduction

Allgemeines zur HL7 Schnittstelle  ACK,  


General Data Used

Drübergehen DTM Format?


Key

Not listed fields are ignored

Default configuration - can be confgured in some extent.

Receiving / Sending seen from custo diagnostic perspective

Common Segments

MSH - Segment


Default Usage
MSH-3

Sending Application. 

Receiving: Expected value can be configured (default empty) in custo diagnostic and works as a filter - meaning messages with deviant content in MSH-3 are ignored.

Sending: Sending Application, e.g. "CUSTO" can be configured in custo diagnostic.

MSH-4

Sending Facility

Receiving: Expected value can be configured in custo diagnostic and works as a filter - meaning messages with deviant content in MSH-4 are ignored. 

Sending: Sending Application, e.g. "CUSTO" can be configured in custo diagnostic. When a tenant system (german "Mandantensystem") is configured, this field contains the information about the tenant. See "Tenant System" below. 

MSH-5

Receiving Application

Receiving: Expected value can be configured (default empty) in custo diagnostic and works as a filter - meaning messages with deviant content in MSH-3 are ignored.

Sending: Receiving Application, can be configured (default empty) in custo diagnostic e.g. "HIS".

MSH-6

Receiving Facility

Receiving:  Expected value can be configured in custo diagnostic and works as a filter - meaning messages with deviant content in MSH-3 are ignored.  When a tenant system (german "Mandantensystem") is configured, this field should contain the information about the tenant. See "Tenant System" below. 

Sending: Receiving Facility, can be configured (default empty) in custo diagnostic e.g. "HIS".

MSH-7Date/Time Of Message, see HL7 Standard.  ##DTM##
MSH-9

Message Type, see HL7 Standard.  By default, custo diagnostic sends Messagetype^Event^Message-Structure, e.g.  ORU^R01^ORU_R01

MSH-10Message Control ID, see HL7 Standard
MSH-12

Version ID, 

Sending: Custo diagnostic supports Version 2.3, 2.5 

##Verhalten bei eingehenden Nachrichten noch prüfen, wenn 2.7  2.9 oder 1.5 o.ä. drin steht##

MSH-13see HL7 Standard
MSH-15##prüfen, sieh auch DS-840)##
MSH-16##prüfen, sieh auch DS-840)##
MSH-18

Character Set  ##zu prüfen##

Receiving: see Standard

Sending: By Default empty, ##Tatsächlich verwendetes CodeSet? ???? is used to encode message## 







PID - Segment


PV1 - Segment



Patient Data Logic (ADT)



Tenant System




Orders (ORM)



Observation Results (ORU)



Physical Transport