The custo diagnostic software version 4.6.4 and higher do support the internal SpO2 and Blood pressure measurement from the Lode custo er2100 treadmill.
There are a couple of things that needs to be fulfilled before this will work. in this item a summary of these requirements.
...
- The treadmill has to be equipped with a SpO2 option and a internal blood pressure option.
- The above options will only work in combination with:
- a programmable control unit.
- an "QRS trigger option" from custo ( this is an option that need to be bought from custo the part number is #12149/LODE)
- minimum custo diagnostic version 4.6.4
- minumum embedded software on the lode custo er2100 treadmill ( the custo diagnostic software will check this as well)
- CIM module 1.0.18
- TCM module 1.0.12
- KIM module 1.0.2
- UIM module 1.0.24
The Lode custo er2100 treadmill should be set to the Lode Treadmill Protocol V2 communication protocol the baud rate should be set to 19.200 baud ( in case you would change this setting you have also change it into the custo diagnostic software)
...
Make sure that you have the custo diagnostic trigger option 12149/LODE and activate the trigger in custo diagnostic.
Make sure you have the trigger activated in the settings of the stress ECG
...