Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

The custo diagnostic software version 4.6.4 and higher do support the internal SpO2 and Blood pressure measurement from the Lode custo er2100 treadmill.

There are a couple of things that needs to be fulfilled before this will work. in this item a summary of these requirements.

...

  • The treadmill has to be equipped with a SpO2 option and a internal blood pressure option.
  • The above options will only work in combination with:
    1.  a programmable control unit.
    2.  an "QRS trigger option" from custo ( this is an option that need to be bought from custo the part number is #12149/LODE) 
    3.  minimum custo diagnostic version 4.6.4
    4. minumum embedded software on the lode custo er2100 treadmill ( the custo diagnostic software will check this as well)   
      1. CIM module 1.0.18
      2. TCM module 1.0.12
      3. KIM module 1.0.2
      4. UIM module 1.0.24

The Lode custo er2100 treadmill should be set to the Lode Treadmill Protocol V2 communication protocol the baud rate should be set to 19.200 baud ( in case you would change this setting you have also change it into the custo diagnostic software)

...

Make sure that you have the custo diagnostic trigger option 12149/LODE and activate the trigger in custo diagnostic.

Image RemovedImage Added


Make sure you have the trigger activated in the settings of the stress ECG

...