custo diagnostic will support the internal SpO2 and Blood pressure measurement devices from custo diagnostic version 4.6.4.

The custo diagnostic software version 4.6.4 and higher do support the internal SpO2 and Blood pressure measurement from the Lode treadmill.

There are a couple of things that needs to be fulfilled before this will work. in this item a summary of these requirements.

The minimum requirements for the treadmill are the following:

  • The treadmill has to be equipped with a SpO2 option and a internal blood pressure option.
  • The above options will only work in combination with:
    1.  a programmable control unit.
    2.  an "QRS trigger option" from custo ( this is an option that need to be bought from custo the part number is #12149/LODE) 
    3.  minimum custo diagnostic version 4.6.4
    4. minumum embedded software on the lode treadmill ( the custo diagnostic software will check this as well)   
      1. CIM module 1.0.18
      2. TCM module 1.0.12
      3. KIM module 1.0.2
      4. UIM module 1.0.24

The Lode treadmill should be set to the Lode Treadmill Protocol V2 communication protocol the baud rate should be set to 19.200 baud ( in case you would change this setting you have also change it into the custo diagnostic software)

In custo diagnostic the following settings should be made. the er2100 V2 should be selected.


Make sure that you have the custo diagnostic trigger option 12149/LODE and activate the trigger in custo diagnostic.


Make sure you have the trigger activated in the settings of the stress ECG