You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 6 Next »

Introduction:

The article discribes the Workflow and the configuration from the file-Interface from certain manufacturers (see list), for cardio pulmonary exercise Testing.

Custo diagnostic is always master and the Spirosystem slave. That means: This spirosystem can only be controlled by custo diagnostic!!

Manufacturer which support this interface

Viasys

SensorMedics

Jäger

CareFusion

Ganshorn

Connection about shared File "ergospir.dat"

diag writes the  parameters every 5 seconds during the spiroergometry exam Text-file c:\\temp\ergospir.dat.

The path is configurable.

Configuration diag

diag > Settings > System (1)> System (2)> Workstation (3) >cardio pulmonary exercise Testing = for example: Ganshorn (LF8.5G) (4)


 

file "ergospir.dat"

diag writes a text file every 5 seconds. with the current ergometry data.

Path

c:\\temp\ergospir.dat

 

The path is configurable in the win.ini file in the Windows directory:

[CAS_RECORD]

SpiroDirectory=C:\\Temp


Workflow for Start from cardio pulmonary exercise Testing :


  1. Start the Spirosystem first and then 
  2. Start custo diagnostic Project: "cardio pulmonary exercise Testing."
  3. select the Patient, Profile and click start 
  4. When switching to the Monitoring page in custo diagnostic, the patient data become now transferred to the Spirosystem
  5. Start the exercise Testing in custo diagnostic