Preconditions

- Firmware custo router 20170303142702 (Image: "custopi-image-all-raspberrypi3-20170303142702.rootfs.rpi-sdimg")

- custo diagnostic 4.5.1

Settings in the Service Center:

Einstellungsname

Defaultwert

Kommentar

Geräte\Devices\VirtualRouterSIO\ActivefalseIf this value is set to true, the processes for the virtual SIO are displayed internally
Geräte\Devices\VirtualRouterSIO\VirtualSIOPort4000The port on which the TCP port handles serial communication
Geräte\Devices\VirtualRouterSIO\VirtualSIORouterHost
Hostname or IP address of the custo routers
Geräte\Devices\VirtualRouterSIO\VirtualSIORouterPort8090Port under which the custo router is reachable.


Device Settings in custo diagnostic

com IR Settings:

If the custo router communication is activated, a "Search Virtual SIO" button will appear on the COM-IR setup page. After pressing the button, all COM-IRs will appear as "Router SIO x" in the list.,


Currently, only one button is searched for an active custo router, the settings retain their validity, even if the router is switched off.

Recorder Settings:

Under ABPM -> connection → Device -> custo screen 100/200/300/400 and as connection com IR select. The corresponding router port is then displayed as hardware


For testing, all recorder information can be read on the recorder page: