Introduction:
The article discribes the Workflow and the configuration from the file-Interface from certain manufacturers (see list), for cardio pulmonary exercise Testing.
Custo diagnostic is always master and the Spirosystem slave. That means: This spirosystem can only be controlled by custo diagnostic!!
Manufacturer which support this interface
Viasys
SensorMedics
Jäger
CareFusion
Ganshorn
Connection about shared File "ergospir.dat"
diag writes the parameters every 5 seconds during the spiroergometry exam Text-file c:\\temp\ergospir.dat.
The path is configurable.
Configuration diag
diag > Settings > System (1)> System (2)> Workstation (3) >cardio pulmonary exercise Testing = for example: Ganshorn (LF8.5G) (4)
file "ergospir.dat"
diag writes a text file every 5 seconds. with the current ergometry data.
Path
c:\\temp\ergospir.dat
The path is configurable in the win.ini file in the Windows directory:
[CAS_RECORD]
SpiroDirectory=C:\\Temp
Workflow for Start from cardio pulmonary exercise Testing :
- Start the Spirosystem first and then
- Start custo diagnostic Project: "cardio pulmonary exercise Testing."
- select the Patient, Profile and click start
- When switching to the Monitoring page in custo diagnostic, the patient data become now transferred to the Spirosystem
- Start the exercise Testing in custo diagnostic